Skip to main navigation Skip to search Skip to main content

A Simple Apparatus for Measuring Cell Settling Velocity

    • Cleveland State University

    Research output: Contribution to journal › Article › peer-review

    Abstract

    Accurate cell settling velocity determination is critical for perfusion culture using a gravity settler for cell retention. We have developed a simple apparatus (a 'settling column') for measuring settling velocity and have validated the procedure with 15-μm polystyrene particles with known physical properties. The measured settling velocity of the polystyrene particles is within 4% of the value obtained using the traditional Stokes' law approach. The settling velocities of three hybridoma cell lines were measured, resulting in up to twofold variation among cell lines, and the values decreased as the cell culture aged. The settling velocities of the nonviable cells were 33-50% less than the corresponding viable cells. The significant variation of settling velocities among cell populations and growth phases confirms the necessity of routine measurement of this property during long-term perfusion culture.

    Original languageAmerican English
    JournalBiotechnology Progress
    Volume26
    DOIs
    StatePublished - Sep 1 2010

    Keywords

    • Cell culture
    • Perfusion culture
    • Gravity settler
    • Cell retention
    • Settling velocity

    Disciplines

    • Biomedical Engineering and Bioengineering