Skip to main navigation Skip to search Skip to main content

An Improved Correction Algorithm for Number Density Measurements Made with the Forward-Scattering Spectrometer Probe

James A. Lock, Edward A. Hovenac

    Research output: Contribution to journalArticlepeer-review

    Abstract

    A correction factor to the number density measured by the Forward Scattering Spectrometer Probe (FSSP) which compensates for dead time and coincidence errors was determined by calculating the probabilities of and the average number of particles in the six possible types of dead time and coincidence events. These probabilities and averages were calculated by means of a probabilistic model based on Poisson statistics. A Monte Carlo computer simulation of the FSSP operation was also carried out and the number density correction factor was compared with the Monte Carlo data. For an actual number density of 2000/cm 3 , it was found that the measured number density was of the order of 300/cm 3 .

    Original languageAmerican English
    JournalREVIEW OF SCIENTIFIC INSTRUMENTS
    Volume60
    DOIs
    StatePublished - Jun 1 1989

    Disciplines

    • Physics

    Cite this